3ω slope comparative method for fluid and powder thermal conductivity measurements
Abstract
By analyzing the relationship among the heat penetration depth, measurement frequency and detector characteristic parameters, a simple and practical 3ω slope comparative method has been proposed. The corresponding measurement system for measuring the thermal properties of fluids and powder materials was established and verified using several specimens with known thermophysical parameters, such as alcohol, distilled water, and air. Compared to the two-dimensional model, the data processing of the method is relatively simple and quick. Due to the elimination of errors introduced by the detector parameter measurement, the measurement accuracy of the method is higher than the conventional one-dimensional model. By using an appropriate frequency range, the new method is time saving and convenient for measuring the thermal conductivity of fluids and powders with low thermal conductivity. Based on the analysis, the effective thermal conductivity of nano-SiO2 powder is accurately determined.