Review of methods to improve the performance of linear array-based photoacoustic tomography
Abstract
As a hybrid imaging modality that combines optical excitation with acoustic detection, photoacoustic tomography (PAT) has become one of the fastest growing biomedical imaging modalities. Among various types of transducer arrays used in a PAT system configuration, the linear array is the most commonly utilized due to its convenience and low-cost. Although linear array-based PAT has been quickly developed within the recent decade, there are still two major challenges that impair the overall performance of the PAT imaging system. The first challenge is that the three-dimensional (3D) imaging capability of a linear array is limited due to its poor elevational resolution. The other challenge is that the geometrical shape of the linear array constrains light illumination. To date, substantial efforts have been made to address the aforementioned challenges. This review will present current technologies for improving the elevation resolution and light delivery of linear array-based PAT systems.
1. Introduction
Photoacoustic tomography (PAT) is a promising biomedical imaging modality that offers high-resolution optical absorption imaging of deep tissue. The hybrid acoustic detection of optical absorption overcomes the optical diffusion limitation.1,2,3,4,5 PAT forms images based on the photoacoustic (PA) effect, which was first discovered by Alexander Bell more than 100 years ago. The research for photoacoustic imaging (PAI) developed very slowly until the 1990s. At this time, the development of high-power pulsed lasers, sensitive ultrasonic transducers, and data acquisition systems triggered the rapid development of PAI. Since then, the technique has evolved rapidly and became one of the most exciting biomedical imaging techniques in this decade. In the last few years, PAT has been demonstrated in various preclinical and clinical applications, such as cardiology imaging,6 oncology imaging,7 vascular biometrics,8 functional imaging,9,10,11,12 breast cancer screening,13,14 and guidance of lymph node biopsy.15
In terms of acoustic signal detection configuration, various ultrasonic transducer arrays, such as ring array,16,17 linear array,18,19,20,21,22 and hemispheric array,13,14 have been implemented in the PAT system to meet the demand of preclinical and clinical applications. Among these, linear transducer arrays are probably the most widely used. As the name implies, the piezoelectric elements of the linear transducer array are arranged in a line, thus, producing a rectangular field of view. Compared to other arrays, linear arrays are handheld operable, convenient, and very affordable. It can be easily manufactured and integrated with light sources and conventional ultrasound systems.
Although linear array-based PAT is promising and has been quickly developed, it still faces challenges that impair the imaging performance of the system. The spatial resolution of a linear transducer array can be represented in three axes. Within the imaging plane shown in Fig. 1(a), the axial direction is along the axis that is perpendicular to the transducer element surface, and the lateral direction is perpendicular to the axial direction. The elevational direction is orthogonal to the rectangular imaging plane.

Fig. 1. Schematic drawing of a linear array transducer. (a) Top: Imaging plane of the linear array transducer. Bottom: Side view of the focal zone of the linear array transducer. (b) Light delivery scheme for side-illumination PAI.
In a linear array-based PAT, the axial resolution (Fig. 1(a)) is primarily defined by the center frequency of the transducer, while the lateral resolution (Fig. 1(a)) is determined by the element pitch of the array. Typically, the axial resolution equals half of the central acoustic wavelength, while the lateral resolution equals one central acoustic wavelength.5
Along the elevation direction, the elements of the linear array transducers are usually cylindrically focused to achieve a cross-sectional imaging plane (Fig. 1(a)). For 3D PAI, the volumetric image can be obtained by stacking two-dimensional (2D) images. In this case, the linear transducer array has to be scanned along the elevation direction. However, because of the fixed cylindrical focus, the resulting volumetric images have a poor spatial resolution along the scanning axis. As shown in Fig. 1(a), the elevation resolution is the highest at the acoustic focus, and it degrades as the object moves away from the focus. Even at the elevational focus, both the axial and lateral resolutions are better than the elevational resolution by at least three times.5,23,24
Besides the nonisotropic spatial resolution, the arrangement of the linear elements limits the light delivery to the image plane of the linear array-based PAT systems. Since the image plane is right under the elements’ surface, the light can only be delivered from the side of the probe for most linear array-based PAT systems (Fig. 1(b)). The side illumination creates additional challenges because it is hard to control the light incident angle according to the distance/depth of the object.
Over the past few years, numerous researchers have made substantial efforts to address these limitations. In this review, current technologies to improve elevation resolution and light delivery of the linear array-based PAT will be presented.
2. The Principle of PAT
Before talking about the details of each system, we first introduce the principle of PAT (Fig. 2).

Fig. 2. Principle of PAI.
PA signals are typically generated by shorted-pulsed lasers. Upon laser irradiation, the object will absorb the photon energy and experience transient thermos-elastic expansion, resulting in the generation of wideband ultrasound signals. For an effective ultrasound signal generation, the pulse duration is typically within a few nanoseconds, which is lower than both the thermal and stress confinements.25 After the wideband emission, the induced acoustic waves are then detected by ultrasound transducers. Finally, the distribution of optical absorption can be obtained by performing an image reconstruction of the detected signals. The PA pressure (p0)p0) can be written as the following equation :
The sound pressure propagates to the tissue surface at different time delays due to the various location of the sound source and properties of the tissue. Linear array-based PAT uses an array of ultrasound transducers to record the sound waves. The distribution of the initial sound pressure or electromagnetic absorption can be obtained by inversely projecting back the detected sound waves. There are many approaches for PAT imaging reconstruction. Among them, the time-domain back-projection reconstruction algorithms are probably the more widely used.26 These methods assume a homogeneous acoustic property in tissue, and the speed of sound in the soft tissue is relatively constant at around 1.5mm/μμs.27
3. Technologies for Improving the Elevation Resolution
To improve the elevational resolution of linear array-based PAT, researchers have proposed numerous solutions. They include new scanning geometries, new detection hardware designs, and advanced image reconstruction algorithms.
3.1. Modifying the scanning geometries to improve the elevation resolution
Gateau et al. first investigated a novel scanning geometry that combined linear and rotational scanning to achieve nearly isotropic 3D spatial resolution.28 As shown in Fig. 3(a), the proposed geometry employs two movements of the linear array, including translation and rotation. The linear array was perpendicularly mounted on a linear translation stage, which was further fixed on a rotary stepper motor so that the stage could be rotated. The entire scanning setup was moved to discrete positions along the contour of a polygon. Each of the translation ranges (polygon sides) was 13.5mm and tangent to a circle centered on the axis of rotation. The entire rotation range was 180°180∘. A 128-element linear array (5.0/7.0MHz, Acuson L7, Siemens) was used to test this method. The spatial resolution of the PAT system was measured through imaging 50mm diameter microspheres. The resulting spatial resolutions were 130μμm in-plane and 330μμm in elevation.

Fig. 3. Different scanning geometries for improving elevation resolution. (a) The rotational scanning geometry proposed in Ref. 28. (b) The translate-tilting scanning geometry proposed in Ref. 29. (c) The bi-directional scanning geometry proposed in Ref. 30. (d) The sample-rotational scanning geometry proposed in Ref. 31.
Upon the initial study of the combined linear and rotational scanning geometry, Gateau et al. proposed a novel single-sided access rotate-translate linear array-based PAT scanner.29 In this system, a 128-element linear array (Vermon, Tours, France) with a 15 MHz center frequency, 101μμm element pitch, and 1.5 mm cylindrical focus was applied. A motorized translation stage and goniometer stage were used to combine translation and rotation motions. The maximum travel range was 50mm for the translation stage and 90°90∘ for the rotation stage. As shown in Fig. 3(b), the linear transducer array was mounted on the goniometer stage and then translated by the translation stage. To ensure that the motion of the elements’ centers was within the XY-plane, the length axis of the linear array was mechanically aligned to match the rotation axis of the goniometer. During the scanning, the array was continuously and simultaneously translated and rotated from −45°−45∘ to 45°45∘. Similar to their initial study, microspheres phantom was used to test the spatial resolution of the PAT system. The spatial resolution results show that the proposed system has a quasi-isotropic 3D resolution of ∼∼170μμm. The improvement of spatial resolution in the translation direction is almost one order of magnitude.
Besides the translate-rotate scanning geometry, Schwarz et al. proposed a bi-directional scan method with two array positions perpendicular to each other to improve the elevation resolution.30 As shown in Fig. 3(c), two linear scans were conducted in perpendicular directions (xx- and zz-direction) for bi-directional scanning. A 24MHz 128 element linear transducer array (LA28.0/128, Vermon, Tours, France) with a 1.5mm focus and 70μμm pitch was first scanned along the elevation direction and was then rotated by 90°90∘ and scanned across the same region again. Numerical simulation and experimental phantom resolution studies both demonstrated that the method yields an improved elevation resolution with only minor losses in lateral resolution.
Li et al. also proposed a combined linear and rotational scanning method to achieve isotropic resolutions for linear array-based PAT.31 Instead of rotating or translating the array over the sample, a rotating sample and translational scanned linear array were adopted, as shown in Fig. 3(d). In each scan step, a commercial linear array with 256 elements (LZ250, 21MHz center frequency, Visualsonics Inc., Canada) was first elevationally scanned. Then the sample was rotated around its center in a 2oo angular step size and scanned again until 180oo have been covered. Carbon fiber was chosen to quantify the elevation resolution of the proposed system, and the resolution result shows that this method improves the elevational resolution by 10 times in comparison with the conventional single elevational scan.
In principle, the various scanning geometries discussed above improve elevation resolution by converting the elevation direction into axial or lateral directions. This conversion overcomes the intrinsic limitation of the linear array transducer. However, such complicated scanning geometry often means a prolonged scanning time. The large data size also takes more time to process for image reconstruction.
3.2. Modifying the detection scheme to improve the elevation resolution
Without modifying the scanning geometry, Wang et al. proposed a fundamentally different approach to improve elevation resolution. The technique is based on acoustic diffraction through a thin slit, which essentially improves the receiving angle of the array along the elevation direction (Fig. 4).32 The imaging system exploited an ATL/Philips L7-4 128-element linear transducer array (5MHz center frequency, 25mm elevation focus) and a thin slit formed by two metal blades with 500μμm thickness. The bottom blade was fixed in position while the top blade was mounted on a translation stage, allowing for easy and precise control of the slit opening (from 300μμm to 1000μμm). The thin slit diffracts the incoming PA waves along the elevation direction and enables 3D image reconstruction. Through phantom and in vivo experiments, the authors demonstrated that the slit improves the elevation resolution by up to 10 times without compromising scanning time.31 Besides improvement in elevation resolution, the slit also improves the signal-to-noise ratio (SNR) due to a larger receiving aperture. Figure 4 displays the vascular image results of a human palm acquired by slit-based linear PAT.

Fig. 4. Slit-based linear PAT proposed in Ref. 33. (a) Schematic drawing of the slit-based linear-array PAT system. (b) 3D palm vascular images acquired by the slit technology. This work by Y. Wang et al., Ref. 33 is licensed under a Creative Commons Attribution 4.0 International License.
3.3. Imaging algorithms for improving the elevation resolution
Other than modifying the system hardware, advanced imaging algorithms can also improve the elevation resolution of the linear array-based PAT system. Wang et al. proposed the integration of two advanced image reconstruction techniques: coherent weighting (CW) and focal-line (FL) 3D image reconstruction.34,35
The CW algorithm calculates the coherence of received PA signals and assigns a weighting factor to the delay-and-summed signals. The value of the coherent weighting factor (CWF) ranges from 0 to 1 and can be applied to the reconstructed image. This factor enhances the coherently summated signals while suppressing the out-of-phase signals and randomized noise. The CW factor is applied in 3D through the FL reconstruction algorithm, which precisely calculates the time of arrival in 3D space. Previous studies have indicated that the FL algorithm can improve the elevation spatial resolution up to the size of the acoustic focus (Fig. 1(b)) regardless of the axial distance of the object.34 The combined CW–FL algorithm was validated through numerical simulation and was experimentally tested in both phantom and human subjects. Both of the experiments’ results proved that the method can significantly improve the elevation resolution by 4 times; moreover, the method also suppresses noise and offers higher SNR.
Alshaya et al. reported another algorithm named the36 Filter Delay Multiply and Sum (FDMAS) beamforming technique, which is an adaptive beamforming technique that depends on the autocorrelation between delayed radio frequency (RF) data. Due to the autocorrelation operation, the FDMAS beamforming technique cannot be applied directly in 3D PAI. Instead, the FDMAS is used to beamform the RF data in the lateral direction and elevation direction separately, and then combine them. This FDMAS beamforming technique not only improved the elevation resolution out of the focal length, but also reduced the clutter signals and background noise of the PA image. The FDMAS beamforming technique was compared with the Delay and Sum (DAS) beamforming technique in terms of spatial resolution and SNR. The improvement in the elevation resolution and SNR are about 30% and 13dB compared to DAS, respectively.
Table 1 summarizes different techniques used to improve the elevation resolution.37 It can be seen that the last three methods possess the fastest speed because they require only a single translational scan, while the first two methods provide the best improvement in elevation resolution because they translate the elevation direction into axial direction during the scan. The middle two methods convert elevation direction into lateral direction during the scan. Depending on the scanning method, the speed ranges from medium to slow.
Name | Scanning method | Speed | Modification of transducer array | Improvement in elevation resolution |
---|---|---|---|---|
Rotational scanning geometry38 | 180°180∘ rotational scan +translational scan at each 1.5°1.5∘ interval | Slow | No | High |
Translate-tilting scanning geometry29 | Continuous and simultaneous rotation and translation scan | Slow | No | High |
Bi-direction scanning geometry30 | Two translational scans perpendicular to each other | Medium | No | Medium |
Sample-rotational scanning geometry31 | The sample is rotated in increments and linearly scanned | Slow | No | Medium |
Slit PAT geometry32 | One translational scan | Fast | Yes (slit) | High |
CW–FL35 | One translational scan | Fast | No | Medium |
FDMAS36 | One translational scan | Fast | No | Medium |
4. Technologies for Improving Light Delivery Efficiency
Due to the linear arrangement and cylindrical focus of the array elements, the imaging plane of linear array transducers is a cross-sectional plane that extended along the axial direction. As we introduced in Sec. 2, the intensity of the PA signal primarily depends on the local light fluence and absorption coefficient. To achieve optimal PAI depth, we need to ensure that the majority of light goes to the imaging plane. In this case, coplanar light delivery and acoustic detection would be the optimal geometry. In reality, due to the geometric shape of linear array transducers, the light is typically delivered from the side of the linear array. Because of the noncoaxial light delivery and acoustic detection, a good amount of the light energy will be wasted outside of the imaging plane. The large incident angle also increases the light travel distance. Both efforts lead to a shallower imaging depth in side-illumination PAT systems. This conclusion is supported by Monte Carlo simulations,39,40 which found that smaller incident angles could significantly improve the light fluence in deep imaging regions.
In this section, we will first discuss technologies for improving the efficiency of side illumination, and then introduce methods to achieve coplanar (coaxial) light delivery and acoustic detection.
4.1. Adjustable side light delivery
Liu et al. designed and built a handheld, real-time PAI system with manually adjustable side light delivery.41 To optimize the laser light delivery, a finite element simulation was employed to evaluate the influence of the incident angle and interval between the two arms of the fiber bundle regarding light fluence. Phantom experimental validations were performed and the results were consistent with the simulations results. These results indicated that the distance between the two arms of the fiber bundle played a major role in light fluence propagation.
Besides the manual adjustment, Sangha et al. reported a motorized adjustable PAT probe.42 The adjustable PAT holder (Fig. 5) consists of an external linear stepper motor and two bipolar stepper motors. This holder can vertically translate the ultrasound transducer and adjust the fiber bundle’s rotation and angle. These adjustments could improve the optical fluence of different tissue layers, thereby increasing the penetration depth and SNR. By tuning the fiber orientation, probe-skin reflection artifacts were reduced and light distribution in the image acquisition plane was improved. They validated the motorized PAT probe through Monte Carlo simulations, ex vivo imaging, and in vivo imaging. The ex vivo results showed several millimeter improvements in penetration depth, while the in vivo results showed a 62% increase in lipid SNR.

Fig. 5. The adjustable probe proposed in Ref. 42. Exploded (a), presentation (b), and constructed (c) images of the probe. Reprinted from Photoacoustics, 12, G. S. Sangha, N. J. Hale, C. J. Goergen, Adjustable photoacoustic tomography probe improves light delivery and image quality, 6–13., Copyright (2018), with permission from Elsevier.
4.2. Coaxial light delivery and acoustic detection
Although side light delivery is widely used in linear array-based PAT systems, coaxial light delivery and acoustic detection is still the optimal light delivery method. The first coaxial system was proposed by Montilla et al., who utilized an optically transparent acoustic reflector to separate light illumination and acoustic detection into two directions.43 As shown in Fig. 6, the glass plate is transparent to light while the acoustic waves were reflected by 90°90∘, achieving coaxial illumination and detection. The performance of the system was demonstrated by imaging a mouse with a pancreatic tumor.

Fig. 6. Cross-sectional view of the photoacoustic enabling device (PED) design.43 In this design, a glass plate was used as a light and sound combiner.
Lee et al. proposed a handheld probe in which the illuminated light beam axis coincides with the axis of the ultrasound through a beam combiner.44 The beam combiner doubly reflects the acoustic waves to the linear array transducer and the transmission losses of the light and acoustic beams are low. Moreover, the laser output was shaped to a line beam by three cylindrical lenses (Fig. 7). In their small animal experiments, the lymph nodes near the skin could be detected without the need for any supplementary material.44

Fig. 7. Schemes of a handheld probe with the illuminated light beam axis coincides with the axis of the ultrasound through a beam combiner.44
Li et al. proposed a similar handheld probe design, utilizing polymethyl methacrylate (PMMA) as an optical/acoustic coupler.27 The optical/acoustic coupler reflects the light beam two times and is permeable to acoustic waves, achieving coaxial light delivery and acoustic detection with the optical fiber bundle and transducer probe parallel to each other. Using the handheld probe, the authors successfully imaged the sentinel lymph node (SLN) in a live rat. Li’s group has also designed various types of compact PA probes to improve the SNR of the system45 and to detect the SLN.46
Recently, Wang et al.47 reported another double-reflector design-based linear optical fiber bundle. Instead of the optics shown in Fig. 7, their system utilized a line output fiber bundle to illuminate the object, so that the whole system is much easier to align. In this design, a compact housing was created to combine both the transducer and fiber bundle output for convenient handheld imaging (Fig. 8). Various phantom and human in vivo experiments were performed to evaluate the efficiency of the system. 3D vascular images of the human forearm were acquired as shown in Figs. 8(b) and 8(c). The images demonstrated that the double-reflector design provides a deeper imaging depth than that of side illumination.

Fig. 8. Schemes of a Double-reflector PAI system.47 (a) The cold mirror is transparent to the near-infrared light, while it reflects acoustic waves by 90°90∘. The second acoustic reflection was achieved by a regular glass. (b) Depth-encoded Maximum amplitude projection (MAP) image of a human forearm, acquired by the side-illumination system. (c) Depth-encoded MAP image of the forearm, acquired by the double-reflector system. White arrows show the same features and the yellow arrow indicates the deepest vessel. This work by Y. Wang et al., Ref. 47 is licensed under a Creative Commons Attribution 4.0 International License.
Table 2 compares different light illumination schemes. It can be seen that while the side illumination is easy to implement, it has limited light penetration depth. The single-reflector design achieves coplanar light illumination and acoustic detection with the minimal wave aberration, however, it is inconvenient for handheld operation. The double-reflector system represents the most compact design among the three. However, it is also associated with a longer acoustic travel distance, which needs to be considered during image reconstruction.47
Light delivery systems | Advantages | Disadvantages |
---|---|---|
Side illumination | Easy to implement. Minimal light and acoustic distortion | Limited imaging depth. |
Single-reflector illumination | Coplanar light illumination and sound detection. | Inconvenient for handheld operation |
Double-reflector illumination | Compact design; Coplanar light illumination and sound detection. | Long acoustic travel distance; potential aberration from two acoustic reflectors. |
5. Summary and Conclusions
Linear array-based PAT is a high resolution, noninvasive, and convenient imaging modality with extensive applications from preclinical laboratory research to clinical patient care. However, due to the unique geometry design of the linear transducer array, it still faces engineering challenges in terms of poor elevation resolution and light delivery that would impair the imaging performance of the PAT system. In this review, we discussed the state-of-the-art of linear array-based PAT systems, introduced the implementation of the imaging system, and elaborated on the reasons for those engineering challenges. We provided a comprehensive review of existing studies and progresses to solve limitations on the elevation resolution and light delivery scheme.
The elevation resolution improvement technologies can be classified into three categories based on their principles. The first category involves new scanning geometries to improve the elevation resolution. Through modified scanning geometries, the elevation direction was converted into axial or lateral directions. However, in consideration of the elongated scanning and imaging reconstruction time, these techniques did not improve the elevation resolution effectively. The second category is based on a brand new technology that improves the elevation resolution by diffracting the acoustic waves through a thin slit. This method provides the highest elevation resolution improvement without modifying the scanning geometry. Furthermore, the slit width can be adjusted to achieve optimal resolution based on the imaging feature. The third category is the simplest method because it improves the elevation resolution purely through imaging reconstruction. Although these algorithm-based methods do not require any hardware modification, their elevation resolution improvement is very limited.
As for the light delivery technologies, there are two major solutions: improving the side-illumination design and achieving coaxial light delivery and acoustic detection. To improve the side-illumination design, it is critical to increase the light fluence within the acoustic detection area. For the first solution, linear array-based PAT systems with adjustable light delivery angles were developed to dynamically control the illumination. As for the second solution, the basic principle is to use acoustic and light combiners to separate or combine light and acoustic beams. The coaxial light delivery and acoustic detection technology has been well demonstrated in imaging animal SLNs and human vasculature. All the aforementioned technologies have greatly improved the imaging capability of the linear array-based PAT system.
To further improve the imaging quality, we envision that future studies will combine the elevation resolution improvement techniques and light delivery schemes to achieve both deeper penetration depth and higher spatial resolution in a linear array. With increasing demand in preclinical and clinical applications of PAT, we anticipate that other areas of linear-array PAT, such as light source and data acquisition system, will also be improved to further reduce the system size and detection sensitivity.48,49 In the next few years, we expect to see a broader application of linear array-based PAT systems.