Processing math: 100%
World Scientific
Skip main navigation

Cookies Notification

We use cookies on this site to enhance your user experience. By continuing to browse the site, you consent to the use of our cookies. Learn More
×

System Upgrade on Tue, May 28th, 2024 at 2am (EDT)

Existing users will be able to log into the site and access content. However, E-commerce and registration of new users may not be available for up to 12 hours.
For online purchase, please visit us again. Contact us at customercare@wspc.com for any enquiries.

Multi-frame speckle reduction in OCT using supercontinuum pumped by noise-like pulses

    https://doi.org/10.1142/S1793545819500093Cited by:4 (Source: Crossref)

    Abstract

    In most coherent imaging modality, speckle noise is a major cause that blurs the boundary of tissues and degrades the image contrast. By utilizing the unique properties of supercontinuum (SC) generated by noise-like pulses (NLPs) and a simple multi-frame averaging technique, we achieved significant speckle reduction in spectral domain optical coherence tomography (SD-OCT). We quantitatively compared the speckle of our proposed method with those of conventional swept source OCT (SS-OCT) and SD-OCT based on commercial light sources. The experimental results show that SC pumped by NLPs combined with noncoherent averaging method achieves better denoising performance in terms of contrast to noise ratio (CNR).

    1. Introduction

    Optical coherence tomography (OCT) is an emerging biomedical imaging modality based on optical interferometric detection, which achieves noninvasive, high-resolution and cross-sectional imaging. However, like other laser imaging modalities, OCT images contain speckles that depend not only on the wavelength and coherent properties of imaging beam but also on the structure of imaging objects.1 Numerous methods have been developed to suppress the speckle noise and improve the image quality. Hardware-based techniques, including spatial,2 angular,3,4 strain,5 and frequency compounding,6,7 can effectively reduce the speckle noise while preserving weak structural signal. On the other hand, these approaches usually require complicated optical design and setup. Software-based techniques, including digital filters,8 wavelet-based method,9 real-time dual window method,10 anisotropic diffusion-based method,11 and probability-based method,12,13 have also been developed rapidly and can achieve speckle suppression effectively. Suitable quantization method may also significantly increase contrast, signal-to-noise ratio (SNR) and visual fineness of the OCT image.14 Nonetheless, most post-processing algorithms require prior knowledge of the imaging object and may suffer from resolution degradation or artifacts.15 In addition, some algorithms are complicated and time consuming,16 incapable of real-time high speed OCT imaging.

    Light source based on supercontinuum (SC) generation provides extremely broad bandwidth and high output power. Thus, SC is particularly suitable for ultrahigh resolution OCT system. SC is usually generated by using ultrashort pulses through nonlinear phenomena in materials, e.g., optical fibers. Noise-like pulses (NLPs), caused by the soliton collapse effect in the lasers,17 have the advantage of broader and smoother spectrum (usually of several tens of nm) compared with conventional mode-locked short pulses.18 In the time domain, the NLP pulse train consists of relatively long (sub-ns), periodic wave packets, which exhibit a fine inner structure of sub-ps pulses with randomly varying amplitude and duration from one wave packet to the next. Typically, the averaged intensity autocorrelation trace of NLPs is double-scaled with a sub-ps peak riding a wide sub-ns pedestal. Since the NLP-based light source is random in nature, its output exhibits low temporal and spatial coherence. This unique property is very helpful to solve the speckle problem of OCT imaging. It has been reported that amplified NLPs could be successfully used for SC generation with standard single-mode fibers.19 In our previous work,20 we have also demonstrated the feasibility of SC pumped by NLPs as the light source for ultrahigh resolution spectral domain OCT (SD-OCT) imaging.

    In this paper, we present a simple multi-frame averaging method for speckle reduction by utilizing the inherent coherence properties of the SC generated by noise-like pulsed fiber laser. Our results demonstrate that this unique light source has very efficient denoising performance in terms of contrast to noise ratio (CNR) and equivalent number of looks (ENL) and cross-correlation (XCOR) value.

    2. Principles and Experimental Setup

    2.1. Experimental setup

    As shown in Fig. 1, we constructed an amplified all-normal dispersion (ANDi) Yb-doped fiber laser capable of generating NLPs. The laser utilized a ring resonator consisting of a 6-m length of Yb-doped fiber, a nonlinear polarization evolution (NPE) port, an optical isolator and a long segment of single mode fibers (SMF28). An optical combiner was used to couple the signal and pumping light together before being sent into the active fiber. The double-cladding gain fiber had a 10-μm core (LIEKKI, Yb1200-10/125DC). Pumping was performed by multi-mode laser diodes (λ=915nm). The NPE-port consisted of a half-wave plate (HWP), quarter-wave plates (QWP) and a polarization beam splitter (PBS).

    Fig. 1.

    Fig. 1. Schematic of the fiber laser system for generating SC. FC: fiber coupler; HWP: half-wave plate; QWP: quarter-wave plate; ISO: optical isolator; MM LD: multi-mode laser diodes.

    The ANDi Yb-doped fiber laser, i.e., the seed laser, generated NLPs with an average power of 200mW (pulse energy of 62.5nJ). The NLPs exhibited a periodic pulse train at 3.2MHz with a spectral bandwidth of 83nm and a double-scaled intensity autocorrelation trace with FWHM of 100ps and 250fs for the pedestal and spike, respectively. Output of the seed laser was coupled into a 2-m length of Yb-doped fiber (LIEKKI, Yb1200-10/125DC) also pumped with a 915-nm laser diode. The amplified NLPs were then used to excite the SC in a 50-m length of standard single mode fiber (POFC, SMF980A). At a pump power of 4.5W, we were able to generate a flat SC spectrum at the central wavelength 1320nm with a spectral bandwidth up to 420nm (in 3-dB bandwidth), as shown in Fig. 2. The average output power reached up to 560mW (pulse energy of 175nJ). The spectral components of SC expanded to longer wavelength at higher pump power, mainly due to cascaded Raman scattering.19

    Fig. 2.

    Fig. 2. The SC spectra generated by the amplified ANDi Yb-doped fiber laser.

    2.2. Imaging processing

    To examine the possible improvement of imaging quality by using the NLP-pumped SC source, some coefficients were used for quantitative comparison. In the previous studies, several blind speckle denoising metrics were used to quantify the imaging quality. They are SNR, CNR, ENL, and XCOR value.

    These coefficients can be defined as follows21:

    SNR=10×log10(mean[I2(i,j)]σ2om),(1)
    CNR=10×log10(μomμbσ2om+σ2b),(2)
    ENLm=μ2omσ2om,(3)
    XCOR=(i,j)Ibefore(i,j)Iafter(i,j)(i,j)[Ibefore(i,j)]2(i,j)[Iafter(i,j)]2,(4)
    where I(i,j) is the intensity at location (i,j) in the selected region of interest (ROI). σb and μb are the standard deviations of the intensity and the mean of the intensity in the background region, respectively; μom is the mean of the pixel values in the mth ROI and σom is the pixel standard deviation. Ibefore and Iafter are the intensity data before and after denoising, respectively; (i,j) are the indices of the images; μafter, σafter, μbefore, and σbefore are the mean and the standard deviation of the intensity in the processed OCT images and original OCT images, respectively. We will discuss in detail the physical meaning of SNR, CNR, ENL, and XCOR later.

    In this experiment, we used a simple sample to demonstrate the imaging quality by applying SC generated by NLPs. As we know, it is relatively easier to see speckle than real signal in the uniform material and the coefficients can be calculated precisely. A sample with more structure would be more intuitive, but how its CNR is calculated is unclear. In order to illustrate our points quantitatively, we used a head cover of fiber connector as the sample, as shown in Fig. 3(a). The corresponding circular-scan two-dimensional (2D) OCT image obtained with a conventional 1310-nm SS-OCT setup is shown in Fig. 3(b). In this image, there are two clear round-shaped boundaries, which indicate the plastic edge of the head cover. Between the two round-shaped boundaries, there are many bright spots. These bright spots apparently result from the speckle noise. Otherwise, because the material of the sample is uniform, there should exist no bright spot and the intensity distribution between two boundaries should be uniform. Because of the sample’s uniformity, it does not matter which area is imaged, even though the three images are quite different. All three systems have different lateral resolution and axial resolution. Besides, each time we switch from one imaging system to another, we cannot ensure to image exactly the same place. Therefore, we finally choose the most stable region for comparison.

    Fig. 3.

    Fig. 3. (a) The picture of fiber connector. The blue dashed box illustrates the circular imaging plane. (b) SS-OCT cross-sectional image of head cover of fiber connector. Scale bar represents 500μm.

    To compare the difference between random noise and speckle in OCT cross-sectional images, a simple averaging method to superimpose the consecutive B-scan images acquired at the same position can be used. The random noise can be reduced in the superimposed image while the speckle noise, as well as the structural signal, will remain with conventional OCT configurations. This is true for typical light sources such as SLED, swept laser sources or femtosecond lasers, because the speckle noise will remain even after image processing of superposition. In contrast, the amplitude, duration and even wavefront of NLP-based light sources are randomly varying from pulse to pulse. It is conceivable that the use of the NLP-pumped SC source may actually induce decorrelation of speckle noise from the sample and thus enable the speckle suppression by the simple method of averaging. Therefore, we utilized it as the light source in a conventional OCT configuration to get OCT images and then measured the denoising metrics of the original and the superimposed images. With the image processing of superposition, the random noise can be reduced and the image quality can be enhanced.

    3. Results and Discussion

    The SD-OCT image obtained by using the NLP-pumped SC light source is shown in Fig. 4. The structure of fiber connector head cover is clearly visible. We may therefore further analyze the images in terms of the denoising metrics. In this study, we actually use six frames and one is shown in Fig. 4(a). Each frame is measured in the same position and under the same conditions. Thereafter, the superposition of the six images is implemented, as shown in Fig. 4(b). Both Figs. 4(a) and 4(b) are shown in linear scale. We calculated SNR, CNR, and ENL by three different ROIs. All ROIs were circled by green boxes in Figs. 4(a) and 4(b). In order to observe the denoising effect more intuitively, we enlarge one pair of green box in Figs. 4(a) and 4(b), then present them in log scale in Figs. 4(c) and 4(d), respectively. In comparison with original images, the superimposed images not only enhance the signal intensity but also reduce the noise level. Furthermore, the intensity distribution between two arc-shaped boundaries is smoother than the previous results as shown in Fig. 3(b), indicating that using the SC generated by NLPs source in OCT systems is helpful to reduce the speckle noise and achieve high quality image.

    Fig. 4.

    Fig. 4. (a) Single frame of head cover using SC pumped by NLPs. (b) Denoised image combined six successive images. The green boxes in these images mark three pairs of homogeneous ROIs used for calculation of SNR, CNR, and ENL, while the blue boxes mark regions of background that were used to obtain SNR and CNR. (c, d) Enlarged and log scaled image in one pair of green box in (a) and (b). Scale bar represents 200μm.

    To evaluate the performance of the light source, we compared three OCT light sources: (1) SC pumped by NLPs using a conventional SD-OCT system, with central wavelength of 1320nm, bandwidth of 420nm, lateral resolution of 10um, and system SNR of 89dB. (2) A commercial SC source (Wuhan Yangtze Soton Laser Co., Ltd.) using the same SD-OCT system, with central wavelength of 1320nm, bandwidth of 300nm, lateral resolution of 10um, and system SNR of 92dB. (3) Commercial swept source (HSL-2000, Santec Corp.) using a conventional SS-OCT system, with central wavelength of 1326nm, bandwidth of 108nm, lateral resolution of 10um, and SNR of 110dB. Original OCT images and corresponding superimposed images are shown in Figs. 46, respectively. It is worth to note that all of the above-mentioned light sources are centered around 1310nm and the OCT setup is independently utilized to get OCT images of the same sample under the same conditions.

    Fig. 5.

    Fig. 5. (a) Single image of head cover using commercial SC. (b) Denoised image combined six successive images. The green boxes mark three pairs of the ROIs, while the blue boxes mark regions of background. (c, d) Enlarged and log scaled image in one pair of green box in (a) and (b). Scale bar represents 200μm.

    Fig. 6.

    Fig. 6. (a) Single image of head cover using commercial SS. (b) Denoised image combined six successive images. The green boxes mark three pairs of the ROIs, while the blue boxes mark regions of background. (c, d) Enlarged and log scaled image in one pair of green box in (a) and (b). Scale bar represents 200μm.

    The imaging quality metrics (SNR, CNR, ENL, XCOR) of single frame and the average of six frames obtained by using the three light sources are calculated and compared with each other, as summarized in Table 1. We choose three different pairs of ROIs to calculate SNR, CNR, and ENL. Their mean values and standard deviations are shown in Table 1. The small standard deviations imply that the choice of ROIs does not affect the results very much. First, SNR measures the ratio between the average amplitude of the intensity in the ROI and the standard deviation in the same region. As shown in Table 1, the improvement of SNR after image superposition is achieved for all the three light sources due to the noise suppression and speckle reduction. Second, CNR determines the contrast between the ROI and the area of background noise. From Table 1, the CNR enhancement of our SC is 3.41.4=2dB, while that of the commercial SC and commercial swept source is 1.8dB and 0.3dB, respectively. Thus for these metrics, the NLP-pumped SC outperforms the other two light sources. Third, ENL measures the smoothness of the ROI which is homogeneous. Therefore, if the image is more homogenous, the coefficient of ENL is larger. The ENL becomes smaller when the homogenous appearance is corrupted by speckle pattern. As we can see, the ENL improvement of the NLP-pumped C is 8.63.4=5.2, while that of the commercial SC and commercial swept source is 4.6 and 0.4, respectively. This indicates that the SC pumped by NLPs enables largest speckle suppression and achieves the best smoothness of ROI in superposition OCT image. Finally, XCOR is an indicator of the similarity between the denoised images and original images. In general, XCOR is usually smaller than 1. But, it is very close to 1 when the denoised image is very similar to the original image. The commercial swept source has its XCOR of 0.985, very close to 1. This means that OCT images with the swept source have very low speckle decorrelation from frame to frame and the denoised image makes little change or improvement compared to the original image. Our custom-built SC has its XCOR of 0.794, indicating fairly good performance. As for the commercial SC, it has its XCOR of 0.944 very close to 1, too. Compared to the original image with the commercial SC, we find that its denoised image after averaging does show some changes and improvement. This is mainly attributed to the reduction of the large intensity fluctuation of the commercial SC after denoising instead of speckle suppression. The large intensity fluctuation increases the mean intensity of the background and hence reduces the difference of mean intensity value between ROI and background region. In the same time, it increases the standard deviation of ROI and background. According to Eq. (2), the numerator is decreased and the denominator is increased, which may lead to (μomμb)σ2om+σ2b less than 1. So in Table 1, we get a negative value of CNR. Furthermore, the improvement of CNR, ENL, and XCOR after denoising with the commercial SC and commercial SS is not as significant as that with our SC. In summary, all light sources show the improvement of SNR using multi-frame averaging. Significantly, SC pumped by NLPs performs better as far as speckle reduction is concerned than other light sources, judging from the CNR, ENL, and XCOR. With the same SD-OCT setup, almost all the imaging quality metrics of our SC are better than the commercial SC source. Especially for the image averaged over six frames, the ENL of our SC is the largest one and the XCOR is the smallest, indicating best smoothness of ROI in OCT image. Therefore, SC pumped by NLPs has very good performance in terms of suppressing speckle noise while preserving image details. This is very important for biomedical imaging.

    Table 1. Quality metrics of speckle suppression for different OCT light sources.

    SC pumped by NLPsCommercial SCCommercial SS
    Single frameDenoisedSingle frameDenoisedSingle frameDenoised
    SNR (dB)6.4 (±0.1)9.7 (±0.3)6.6 (±0.1)9.6 (±0.4)5.6 (±0.1)6.0 (±0.1)
    CNR (dB)1.4 (±0.1)3.4 (±0.2)1.1 (±0.2)0.7 (±0.3)2.0 (±0.1)2.3 (±0.1)
    ENL3.4 (±0.1)8.6 (±0.6)3.6 (±0.1)8.2 (±0.8)2.6 (±0.1)3.0 (±0.1)
    XCOR0.7940.9440.985

    4. Conclusion

    To conclude, we have demonstrated that the unique properties of SC pumped by NLPs can be combined with simple multi-frame compounding method for OCT speckle reduction. Compared with conventional light sources, our approach is more favorable for the decorrelation of successive B-scans and the speckle suppression in terms of image quality metrics of SNR, CNR, ENL, and XCOR value. Because our custom-built SC light source is based on fiber technology, the system is stable, compact, relatively inexpensive, and easy to connect with OCT system. We believe that it will be very promising to be used as a new light source for ultrahigh resolution OCT imaging and other applications as well.

    Conflict of Interest

    The authors have no relevant financial interests in this article and no potential confilicts of interest to disclose.

    Acknowledgments

    This research was supported by the National Natural Science Foundation of China (61227807, 61575108 and 61505034), Tsinghua Initiative Scientific Research Program (2013THZ02-3). C. L. Pan and Y. J. You were supported in part by the Ministry of Science and Technology, Taiwan, under Grant 103-2622-E-007-006-CC2 and by the National Tsing Hua University Research Program Grant 104N2711E1. C.L. Pan is also supported by the Air Force Office of Scientific Research FA2386-13-1-4086. The first three authors contribute equally to this paper.